Optical emission spectrometer is a CCD based spectrometer with a wavelength range of 130 nm to 800 nm. Floor standing and bench top optical emission spectrometer are the two types available. Automatic light path correction, optical system automatically scans the spectral lines to ensure the correctness of received lines and avoid tedious scanning of wave peaks.
Applications
Used for metal and trace elemental analysis, foundry, mechanical engineering, scientific research, product inspection, automobile, petrochemical engineering, shipbuilding, electrics, aerospace, nuclear power, metallic and nonferrous metallic smelting, processing and recycling.
Specifications
| Wavelength range | 130 nm ~ 800 nm |
| Degree of vacuum | Auto control within 6 pa ̴ 20 pa |
| Pixel resolution | 30 pm |
| Grating Focus | 400 nm |
| Grating line | 2400 m1 / mm |
| First order spectral line dispersion rare | 1.2 nm / mm |
| Average resolution ratio | 10 pm / pixel |
| Spark frequency | 100 Hz ̴ 1000 HZ |
| Plasma current | 1 A ̴ 80 A |
| Ignition voltage | > 15000 V |
| Storage temperature | 0 °C - 45 °C |
| Operating temperature | 10 °C - 30 °C, 23 ± 2 °C |
| Power | AC 220 V / 50 Hz |
| Power consumption | Excitation : 700 W Stand by : 100 W |
| Argon quality | 99.999 %, Argon pressure > 4 Mpa |
| Argon consumption | 5L / min during spark mode |
| Dimension | 800 x 700 x 470 mm |
Features
CCD-based, spark discharge OES
Highly integrated and direct reading system
Digital light for ultra-high resolution and high-stability output rate of plasma
Flexible sample clamp design and on-site analysis of samples
The analytical base is Fe base
Customizable bases of Al, Cu, Zn, Fe and Cu, Fe, Cu and Al are available
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.