Introduction

Cyto-Low Speed Centrifuge is a benchtop centrifuge equipped with advanced features for accurate and regulated centrifugation. Our centrifuge accommodates up to 6 tubes, each holding capacity of 30 mL of biological sample for processing and analysis. Enhanced with digital display provides real-time monitoring of centrifugation parameters, allowing users to monitor and adjust settings easily. Upgraded with microprocessor-controlled system improves sample processing accuracy and reproducibility. It utilizes high-speed rotation to separate components of different densities, widely used in biological, medical, and chemical laboratories. The unit is equipped with safety protection systems to ensure operator safety during high-speed operation.

Applications

Cyto-Low Speed Centrifuge is widely used in the field of biochemistry, in separation of blood components, for analysis in diagnostic laboratories and research institutes.

Specifications

Maximum Capacity 6 × 9 mL
Maximum Speed 4000 Rpm
Maximum 2170 × g
Time Range 0 to 99 min
Speed Accuracy ±20r/min
Noise ≤ 55 dB
Power Supply AC220V,50Hz
/ Convert Yes
Dimension 540 × 370 ×280 mm
Weight 28 Kg

Features

Equipped with emergency lock which helps in opening the lid when power goes off

Consists of less noisy brushless DC motor which ensures easy operation

Designed with three layers of protective steel covering and stainless-steel chamber

Light in weight and compact

Automatic lid lock closes unit effortlessly

FAQs

1. What is the maximum sample volume that can be processed in the centrifuge?

The maximum sample volume depends on the type of rotor used with the centrifuge. Typical rotors can accommodate sample volumes ranging from 0.2 mL (microcentrifuge tubes) to 50 mL (conical tubes) or larger for specialized rotors. It is important to check the centrifuge's specifications and rotor compatibility to ensure that the sample volume does not exceed the recommended limits, which could affect centrifugation efficiency and safety.

2. How to balance the centrifuge before operation?

Proper balancing is critical for safe and effective centrifuge operation: 1) Ensure that all sample tubes have the same volume and weight; 2) Place the tubes symmetrically in the rotor, opposite each other; 3) For odd numbers of samples, use a balance tube filled with the same volume of appropriate solution; 4) Check that the rotor is properly seated and secured in the centrifuge; 5) Avoid overloading the rotor beyond its maximum capacity.

3. What is the difference between relative centrifugal force (RCF) and revolutions per minute (RPM)?

RCF is a measure of the centrifugal force applied to the sample, expressed in multiples of gravity (x g), and depends on both the RPM and the radius of the rotor. RPM is a measure of the rotor's rotation speed in revolutions per minute. The relationship between RCF and RPM is given by the formula: RCF = 1.118 × 10^-5 × r × RPM² (where r is the rotor radius in centimeters). For most applications, RCF is the preferred parameter as it directly relates to the separation efficiency, regardless of rotor size.

4. How to maintain the centrifuge to ensure long service life?

Regular maintenance includes: 1) Cleaning the rotor and chamber after each use to remove sample spills and debris; 2) Inspecting the rotor for cracks, corrosion, or wear before each use; 3) Lubricating moving parts according to the manufacturer's recommendations; 4) Calibrating the speed and temperature controls annually or after major service; 5) Ensuring proper ventilation around the centrifuge to prevent overheating; 6) Storing the centrifuge in a clean, dry environment when not in use.

5. What safety precautions should be taken when using the centrifuge?

Important safety precautions include: 1) Always balance the sample tubes properly before operation; 2) Ensure the rotor is correctly installed and secured; 3) Do not open the centrifuge lid while the rotor is still spinning; 4) Use appropriate personal protective equipment (gloves, goggles) when handling samples and cleaning the equipment; 5) Avoid overloading the centrifuge beyond its maximum capacity; 6) Regularly inspect the centrifuge for any signs of damage or malfunction and discontinue use if any issues are found.

Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.