Introduction

Atomic Absorption Spectrophotometer features a 190 to 900 nm wavelength range with ±0.2 nm accuracy. It supports both flame and graphite furnace atomization with fast mode switching. It is fully automatic and PC-controlled, ensuring stable and reliable performance. The system is equipped with a PID-controlled graphite furnace and an 85-position high-precision autosampler for consistent results.

Applications

Atomic Absorption Spectrophotometer is used for trace and heavy metal elemental analysis in food, pharmaceuticals, environmental, and material samples. It is used in research labs, quality control laboratories, and environmental monitoring facilities.

Specifications

Wavelength Coverage190 to 900 nm
Lamp Configuration66 hollow cathode lamps
Wavelength Repeatability±0.1 nm
Wavelength Accuracy±0.2 nm (full wavelength range)
Absorbance Range0 to 2.5 ABS
Raster / Grating1800 lines/mm
Spectral Bandwidth Levels0.1, 0.2, 0.4, 0.7, 1.0, 2.0 nm
Optical Resolution0.2 nm bandwidth separates Mn double lines (279.5 / 279.8 nm), valley-peak energy ratio < 20%
Static Baseline Drift≤ 0.002 ABS / 30 min (Cu)
Background CorrectionBackground signal of 1 ABS reduced by ≥ 50 times (D₂ Lamp)
Flame Atomization ModeAir-acetylene flame
Flame Burner Head Length100 mm
Flame Dynamic Baseline Drift≤ 0.004 ABS / 30 min (Cu)
Flame Sensitivity≤ 0.025 μg/mL / 1% (Cu)
Flame Detection Limit≤ 0.004 μg/mL (Cu)
Flame Precision≤ 0.5% (Cu, absorbance > 0.8 ABS) (RSD)
Flame Safety ProtectionInadequate pressure, power failure, abnormal flame-out, automatic gas cut-off, burner mismatch protection
Flame SensorHigh-sensitivity flame sensor
Graphite Furnace Working TemperatureRoom temperature to 3000 °C
Maximum Heating Rate≥ 2000 °C/s
Temperature Control MethodPID temperature control
Graphite Furnace Characteristic MassCd ≤ 1 pg; Cu ≤ 10 pg
Graphite Furnace PrecisionCd ≤ 3%; Cu ≤ 3% (RSD)
Graphite Furnace SafetyOvercurrent protection
Gas Pressure AlarmAutomatic alarm and stop heating
Cooling Water Flow AlarmAutomatic alarm and stop heating
Graphite Furnace Visual SystemCCD camera real-time monitoring inside graphite tube
Visual Monitoring FunctionsCapillary needle depth, platform position, drying/ashing/atomization process
Autosampler TypeAutomatic sampler
Autosampler Capacity85 positions
Autosampler FunctionsAutomatic injection, automatic dilution, automatic standard curve configuration
Injection AccuracyHigh-precision injection
Flame & Furnace SwitchingMode switching within 10 seconds
Overall Safety SystemComprehensive operator protection
Graphite Furnace Power Consumption7000 W
Graphite Furnace Power Supply220 V AC
Graphite Furnace Dimensions550 × 450 × 280 mm
Host Dimensions500 × 450 × 430 mm
Graphite Furnace Weight60 kg
Host Weight80 kg

Features

Fast mode switching without mechanical adjustment

Stable optical-mechanical design

Real-time graphite furnace visual monitoring

Advanced safety with automatic gas shut-off

Intelligent software with method & curve storage

FAQ

Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.

Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.

Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.

Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.

Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.

Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.