Introduction
Electrolyte Analyzer supports whole blood and urine testing with a 160 µL sample volume, ensuring accurate and efficient electrolyte analysis. It utilizes the Ion Selective Electrode (ISE) method, providing accurate and ion-specific analysis. It features an integrated pilot lamp, ensuring clear indication of sample position and reagent liquid levels. analyzer has colourful touchscreen display, enabling precise monitoring and control.
Applications
Electrolyte Analyzer is used for accurate measurement of key electrolytes in whole blood and urine. It is widely utilized in hospitals, clinical laboratories, emergency units, and biochemical research facilities.
Specifications
| Test Parameters | K⁺, Na⁺, Cl⁻, Ca²⁺, pH, TCO₂ |
| Calculated Parameters | nCa, TCa, AG |
| Measuring Method | Ion Selective Electrode Method |
| Blood Measuring Range | K⁺ - 0.50 to 15.00 mmol/L Na⁺ - 30.0 to 200.0 mmol/L Cl⁻ - 30.0 to 200.0 mmol/L Ca²⁺ - 0.10 to 5.00 mmol/L pH - 4.0 to 9.0 mmol/L TCO₂ - 4.0 to 70.0 mmol/L |
| Blood Accuracy | K⁺, Na⁺, Cl⁻: ≤ ±3.0% Ca²⁺: ≤ ±5.0% or 0.05 mmol/L pH: ≤ 0.15 TCO₂: ≤ 3.0 mmol/L |
| Blood Precision | K⁺, Na⁺, Cl⁻: ≤ 1.5% Ca²⁺: ≤ 1.5% pH: ≤ 1.0% TCO₂: ≤ 3.0% |
| Urine Measuring Range | K⁺ - 5.0 to 150.0 mmol/L Na⁺ - 30.0 to 200.0 mmol/L Cl⁻ - 10.0 to 300.0 mmol/L |
| Urine Accuracy | K⁺, Na⁺, Cl⁻: ≤ ±3.0% Ca²⁺: ≤ ±5.0% or 0.05 mmol/L pH: ≤ 0.15 TCO₂: ≤ 3.0 mmol/L |
| Urine Precision | K⁺, Na⁺, Cl⁻: ≤ 1.5% Ca²⁺: ≤ 1.5% pH: ≤ 1.0% TCO₂: ≤ 3.0% |
| Measuring Time | ≤100 seconds |
| Sample Capacity | 160 µL |
| Sample Type | Whole Blood and Urine |
| Sampling Method | Automatic |
| Operating Temperature Range | 15 °C to 30 °C |
| Humidity Level | ≤85% |
| Display | Colorful touchscreen display |
| Data Storage | Yes |
| Power Consumption | 50 W |
| Power Supply | AC 220V ± 22 V or AC 110V ± 11 V /, 50-60Hz |
| Product Dimension | 408 × 390 × 230 mm |
| Weight | 8 kg |
Features
Sample pump calibration
Automatic sampling
Air bubble detection
Automatic flushing
Solenoid valve system
User-friendly interface
Automatic calibration
Data storage capability
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.