Introduction

Electrolyte Analyzer supports whole blood and urine testing with a 160 µL sample volume, ensuring accurate and efficient electrolyte analysis. It utilizes the Ion Selective Electrode (ISE) method, providing accurate and ion-specific analysis. It features an integrated pilot lamp, ensuring clear indication of sample position and reagent liquid levels. analyzer has colourful touchscreen display, enabling precise monitoring and control. 

Applications

Electrolyte Analyzer is used for accurate measurement of key electrolytes in whole blood and urine. It is widely utilized in hospitals, clinical laboratories, emergency units, and biochemical research facilities.

Specifications

Test ParametersK⁺, Na⁺, Cl⁻, Ca²⁺, pH, TCO₂
Calculated ParametersnCa, TCa, AG
Measuring MethodIon Selective Electrode Method
Blood Measuring RangeK⁺ - 0.50 to 15.00 mmol/L
Na⁺ - 30.0 to 200.0 mmol/L
Cl⁻ - 30.0 to 200.0 mmol/L
Ca²⁺ - 0.10 to 5.00 mmol/L
pH - 4.0 to 9.0 mmol/L
TCO₂ - 4.0 to 70.0 mmol/L
Blood AccuracyK⁺, Na⁺, Cl⁻: ≤ ±3.0%
Ca²⁺: ≤ ±5.0% or 0.05 mmol/L
pH: ≤ 0.15
TCO₂: ≤ 3.0 mmol/L
Blood PrecisionK⁺, Na⁺, Cl⁻: ≤ 1.5%
Ca²⁺: ≤ 1.5%
pH: ≤ 1.0%
TCO₂: ≤ 3.0%
Urine Measuring RangeK⁺ - 5.0 to 150.0 mmol/L
Na⁺ - 30.0 to 200.0 mmol/L
Cl⁻ - 10.0 to 300.0 mmol/L
Urine AccuracyK⁺, Na⁺, Cl⁻: ≤ ±3.0%
Ca²⁺: ≤ ±5.0% or 0.05 mmol/L
pH: ≤ 0.15
TCO₂: ≤ 3.0 mmol/L
Urine PrecisionK⁺, Na⁺, Cl⁻: ≤ 1.5%
Ca²⁺: ≤ 1.5%
pH: ≤ 1.0%
TCO₂: ≤ 3.0%
Measuring Time≤100 seconds
Sample Capacity160 µL
Sample TypeWhole Blood and Urine
Sampling MethodAutomatic
Operating Temperature Range15 °C to 30 °C
Humidity Level≤85%
DisplayColorful touchscreen display
Data StorageYes
Power Consumption50 W
Power SupplyAC 220V ± 22 V or AC 110V ± 11 V /, 50-60Hz
Product Dimension408 × 390 × 230 mm
Weight8 kg

Features

Sample pump calibration

Automatic sampling

Air bubble detection

Automatic flushing

Solenoid valve system

User-friendly interface

Automatic calibration

Data storage capability

FAQ

Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.

Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.

Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.

Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.

Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.

Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.