Introduction
Ion Chromatography stands out with its innovative design and advanced functionalities. It incorporates built-in circulating 3D constant temperature technology, guaranteeing quick temperature stability and accuracy of test data. The chromatograph provides high efficiency and large capacity for diverse ion compositions. Additionally, the self-regenerating electrolytic micro-membrane suppressor and the auto-range conductivity detector contribute to its superior performance. Enables direct detection from ppb to ppm without the need for range adjustments.
Applications
Ion Chromatography is widely used across chemicals, pharmaceuticals, environmental protection, pharmaceuticals, etc.
Specifications
| IC Pump Type | High-pressure and low-pulse two-piston tandem advection pump |
| IC Pump Maximum Pressure | 35 MPa (PEEK) |
| IC Pump Pressure Display Accuracy | ≤ 0.1Mpa |
| IC Pump Flow Stability | (0.2 to 0.5) mL/min ≤3 % (0.5 to 1.0) mL/min ≤2 % >1.0mL/min ≤ 2 % |
| IC Pump Allowable Deviation of Flow | (0.2 to 0.5) mL/min , ±5% (0.5 to 1.0) mL/min, ±3% >1.0 mL/min, ±2% |
| IC Pump Flow Range | 0.001 to 9.999 mL/min |
| IC Pump Pressure Pulse | ≤ 0.5% |
| IC Pump Flow Precision | ≤ 0.1% |
| IC Pump Flow Accuracy | ≤ 0.1% |
| Manual Sample Injector Maximum Pressure | 35 Mpa |
| Manual Sample Injector External Diameter of the Tube | 1/16" |
| Manual Sample Injector Contact Material of the Rotor | PEEK |
| Manual Sample Injector Contact Material of Medium | PEEK / Ceramics |
| Column Heater Operating Temperature Range | 20 to 60℃ (68 to 140℉) |
| Column Heater Controlling Temperature Accuracy | ± 0.01℃ |
| Column Heater Allowable Deviation of Column Heater’s Temperature | ± 1℃ |
| Column Heater Temperature Stability | ≤ 0.05℃/h |
| Conduction Detection System Type | Constant temperature auto-range conductivity detector |
| Conduction Detection System Cell Volume | ≤ 0.8 μL |
| Conduction Detection System Detection Range | 0 to 35000 μS/cm |
| Conduction Detection System Detection Resolution | ≤ 0.0020 nS/cm |
| Conduction Detection System Output Voltage | -6000 to +6000 mv (adjustable) |
| Conduction Detection System Electronic Noise | 0.02 nS |
| Conduction Detection System Baseline Noise | ≤ 0.001 μS/cm |
| Conduction Detection System Baseline Drift | ≤ 0.02μS |
| Conduction Detection System operating Temperature Range | Room temperature +5℃ to 60℃ (41 to 140℉) |
| Conduction Detection System Controlling Temperature Accuracy | ± 0.01℃ |
| Conduction Detection System Temperature Compensation | 1.7 % /℃ |
| Conduction Detection System Maximum Pressure | 10.0 Mpa |
| Conduction detection System Linear Range | ≥ 103 |
| Conduction Detection System Instrument Linearity | ≥ 0.999 |
| Conduction Detection System Quantitative Repeatability | ≤ 1.0% |
| Conduction Detection System Qualitative Repeatability | ≤ 0.5% |
| Conduction Detection System Minimum Detectable Concentration | Cl-≤ 0.0005 ug/mL Li+≤ 0.001 ug/mL BrO₃ -≤0.001 ug/mL |
| Flow System Plastic Flow Path | Made of PEEK materials |
| Flow System Six-way Valve | PEEK material, pressure 5000psi; Independent automatic collecting and flow function |
| Suppressor Type | Self-Regenerating electrolytic micro-membrane suppressor |
| Suppressor Maximum Pressure | 6.0 Mpa |
| Suppressor Dead Volume | <50 μL |
| Power Supply | 220V 50/60 Hz |
| Power | 150 W |
| Exterior dimensions (L × W × H) | 325 × 400 × 500 mm |
| Net Weight | 21 Kg |
| Gross Weight | 28 Kg |
Features
Temperature stability time lesser than 30 minutes
Efficient separation and detection
Analyzes ions with varied compositions
Great durability
Enhances signal responsiveness and accuracy
Ion Chromatography tool is user-friendly
Comprehensive Data Management
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.
Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.