Introduction
Automatic Slide Stainer features three staining racks, each capable of holding 20 slides, and supports an infinite loop working mode. Its dyeing rack features an automated identification function. This stainer supports up to 10 programmable staining protocols and features a multi-functional 18-step workflow. It includes features like priority reagent processing for differentiation and eosin tanks. Our stainer is designed for rapid and reliable HE staining of intraoperative frozen tissue specimens.
This slide stainer is designed to meet the rigorous requirements of modern laboratory operations, ensuring high precision, reliability, and ease of use for various experimental applications.Applications
Our Automatic Slide Stainer is used in surgical pathology and oncology for rapid intraoperative tissue analysis. It is also applied in research labs and transplant medicine for quick evaluation of frozen tissue samples.
Specifications
| Slide Capacity per Rack | 20 slides |
| Staining Time | 4 minutes adjustable |
| Number of Staining Steps | 18 |
| Number of Liquid Cylinders | 15 |
| Number of Programs | 10 programmable methods |
| Working Mode | Infinite loop with 3 staining racks |
| Tank Configuration | 1 hematoxylin (heated), 1 eosin, 1 differentiation, 1 anti-blue, 5 gradient alcohols, 2 xylene cylinders |
| Hematoxylin Heating | Water bath heating with ultrasonic wave, adjustable from 0 to 99°C |
| Temperature Control | Adjustable 0 to 99°C for hematoxylin and blue return tanks |
| Cylinder Processing Time | Adjustable from 1 sec to 59 min per cylinder |
| Return-to-Blue Options | Three methods, including warm water (temperature adjustable) |
| End-of-Process Alert | Yes, with alarm function |
| Dimensions | 810×430×450 mm |
| Packaging Dimensions | 880 × 510 × 535 mm |
| Net Weight | 55 kg |
| Gross Weight | 70 kg |
Features
Computerized water inlet and drainage system
Audible program completion alarm
Programmable heating for water tanks
Efficient and accurate staining
Bench-top design
FAQs
1. What is the complete staining and processing workflow for frozen section specimens using Automatic Slide Stainer ?
The frozen section specimen workflow begins with fixing the specimen in cylinder number 1 for 20 seconds, followed by a quick wash in cylinder number 4 for 10 seconds. Next, the specimen undergoes hematoxylin and eosin (HE) staining in jar number 2 for 1 minute and 20 seconds. This is followed by repeated washing steps in tank number 4 and differentiation in tank number 3 to enhance contrast. The process continues with unbluing in tank number 7 and additional washes to prepare the tissue for eosin staining in tank number 8. Subsequently, a series of alcohol gradient washes in tanks 9 through 13 progressively dehydrate the specimen, ending with clearing in xylene tanks 14 and 15. Each step is precisely timed, ranging from a few seconds to over a minute, ensuring thorough staining, differentiation, dehydration, and clearing for optimal microscopic evaluation.
2. What are the key features that improve usability and workflow in the Automatic Slide Stainer ?
Designed with a bench-top form factor, the Automatic Slide Stainer fits conveniently in most laboratory settings without occupying excessive space. It has an automatic dyeing rack identification system to prevent program errors and ensure correct protocol execution. An audible program completion alarm notifies operators when staining is finished, enhancing workflow efficiency. The device’s automated water inlet and drainage system minimizes manual intervention, streamlining routine operations.
3. In which clinical and research fields is the Automatic Slide Stainer most effectively used?
Automatic Slide Stainer is primarily used in surgical pathology and oncology for rapid intraoperative frozen tissue analysis, aiding timely surgical decisions. It also finds applications in research laboratories for studying frozen tissue samples and in transplant medicine for fast tissue evaluation during organ transplantation procedures. Its fast-staining cycle and accuracy make it valuable wherever frozen section analysis is critical.
4. What staining capacity and workflow does the Automatic Slide Stainer support?
Automatic Slide Stainer features three staining racks, each holding up to 20 slides, allowing simultaneous batch processing of 60 slides. It supports an infinite loop working mode to ensure continuous staining without interruption. It offers a multi-functional 18-step workflow, enabling flexible and programmable staining sequences that can be customized to meet diverse laboratory needs. This makes it ideal for high-throughput pathology labs requiring efficient slide processing.
5. How does the Automatic Slide Stainer ensure staining accuracy and reagent efficiency?
Automatic Slide Stainer stainer incorporates priority reagent processing, especially for critical tanks like differentiation and eosin, ensuring optimal reagent timing and usage. The hematoxylin tank includes ultrasonic water bath heating with adjustable temperatures from 0 to 99°C to enhance nuclear staining quality. Additionally, the device features computerized water inlet and drainage systems, reducing waste and maintaining reagent purity. These features combine to deliver reliable and reproducible staining results.
Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.