Introduction
Automated Atmospheric Distillation Tester determine boiling point of liquid storing up to 5000 results. It ensures easy operation with its ergonomic design and 12" color touch screen. Integrate Intelligent temperature control to optimizes heating for standardized speed. Equipped with efficient Peltier cooling technology that reduces maintenance costs. This instrument is ideal for petroleum testing, providing accurate and reliable results.
This petroleum testing is designed to meet the rigorous requirements of modern laboratory operations, ensuring high precision, reliability, and ease of use for various experimental applications.Applications
Automated Atmospheric Distillation Tester is ideal in petroleum industry, chemical manufacturing, and quality control laboratories, providing accurate fractionation analysis for various liquid samples and applications.
Specifications
| Test Range | 0 to 400 °C |
| Accuracy | ±0.1 °C |
| Receiving Chamber | 0 to 60 °C (0 °C achievable when AT < 15 °C) |
| Condenser Bath | Peltier cooling |
| Condenser Temperature Range | 0 to 60 °C (0 °C achievable when AT < 25 °C) |
| Heating System | Low mass/low voltage heater Unique heating optimization system for fully automatic initial heat settings and heating regulation |
| Safety Features | Monitor the whole test process automatically: Alarm and stop the test automatically if cannot be detected in specified time. Alarm and stop the test automatically if the flow rate exceeds 15mL/min or no change for a long time during test. Alarm and stop heating if the temperature of condenser tube or receiving chamber is higher than 70℃. Built-in fire extinguisher, extinguish fire manually or automatically in case of fire (N2 supply required from lab) The system will turn off if expected dry point cannot be observed |
| Detection | Infrared photoelectric detection |
| Dry Point Detection | Automatic dry point kit (optional) |
| Sample Volume Detection | Stepper motor with high-precision photoelectric sensor |
| Volume Range | 0 to 100 mL, 99 recovered volume points |
| Volume Accuracy | 0.1 mL |
| Pressure Correction | Automatic barometric correction |
| Data Storage | ≥ 5000 results |
| User Interface | 12 |
| I/O Ports | ×2, RS232 ×1, ×1 |
| Data Export | or Printer |
| Standards | D86, 3405, IP 123, GB/T 6536 |
| Operating Conditions | 5℃ to 35℃ 20 to 90%RH No air flow |
| Power Consumption | 2000 W |
| Power Supply | AC 220 V ±10%, 50/60 Hz |
| Dimensions | 525×590×585 mm |
| Packaging Dimensions | 680 × 760 × 930 mm |
| Net weight | 90 kg |
| Gross weight | 100 kg |
Features
High-Precision Level Measurement
Built-in Fire Safety
Self-Diagnosis Feature
Data Storage Capacity
Customizable Testing Methods
FAQs
1. How does the Automated Atmospheric Distillation Tester detect the initial boiling point and manage endpoint accuracy?
The Automated Atmospheric Distillation Tester uses an infrared sensor to spot the first drop of liquid, which marks the Initial Boiling Point (). At the same time, it tracks the amount of liquid collected using a motor that moves the receiver with high accuracy. The final boiling point is worked out by checking how much liquid is collected and comparing it with the test settings. If needed, users can add a dry point kit to get more detail at the end of the distillation process.
2. How does the stepper motor–driven receiver system in the Automated Atmospheric Distillation Tester improve volume measurement?
Automated Atmospheric Distillation Tester uses a motor to move the receiver cylinder, which measures liquid volume in small steps of about 0.1?mL. This helps the system track the distillation process more accurately, especially at key points like 10%, 50%, and 90% recovery. The receiver moves automatically, so no manual changes are needed, which reduces user mistakes. This is very useful when testing fuel products that must meet strict rules.
3. What real-time parameters are displayed during the distillation run in the Automated Atmospheric Distillation Tester , and why are they importa
Automated Atmospheric Distillation Tester shows important values on the screen during the test like vapor temperature, heater temperature, receiver temperature, recovered volume, time, and distillation speed. These values help the user see how the sample is behaving and catch any problems early. Watching these values in real time lets users stay in control and make fast decisions if something goes wrong. This is helpful in both research labs and regular product testing.
4. How does the Automated Atmospheric Distillation Tester ’s real-time distillation curve help during testing?
The Automated Atmospheric Distillation Tester shows a live curve on the screen that matches temperature with how much liquid has been collected. This helps users clearly see how the sample is performing during the test. If the curve changes too fast or looks strange, it might mean there's a problem with the sample or the setup. Seeing this right away helps users fix issues quickly and understand the sample better.
5. What should users do if unexpected results or sample behaviour is observed in Automated Atmospheric Distillation Tester ?
Automated Atmospheric Distillation Tester gives warning messages and a graph that helps users understand what went wrong if the results don’t look right. Some reasons could be dirty glassware, wrong settings, or a bad sample. The system makes it easy to stop the test, clean parts, change the setup, or run the test again. This helps save time and keeps the lab running smoothly.
Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.