Introduction

comes with high precision syringe sampling and semiconductor laser sensor. Designed using photoresist method principle, it detects size and number of solid particles in a liquid sample. It’s a high quality performance instrument, designed in accordance with international standards such as ISO4406:1999. This analytical equipment is designed to meet the rigorous requirements of modern laboratory operations, ensuring high precision, reliability, and ease of use for various experimental applications.

Applications

Used in petroleum, metallurgy, automobile, aviation, chemical and other fields to detect insoluble particles present in hydraulic oil, petroleum oils, turbine and insulating oils, liquid polymer solution.

Specifications

Measurement range
1 to 250 µm (25 sensors)
1 µm to 320 µm (32 sensors)
2 µm to 400 µm (40 sensors)
Light Source Semiconductor laser light
Measurement Channels 8 channels
Particle size 1 µm (2 µm) to arbitrarily set within 100 µm
Sensitivity
1 µm (ISO04402)
4 µm (C) (ISO11171) (25, 32 sensor)
2 µm (ISO4402) (40 sensor)
Sample volume ≥ 0.5 ml
Sample accuracy Better than ± 1 %
Sampling rate 5 ml/ min to 60 ml/ min
Resolution < 10 %
Coincidence error limit 10,000/ ml (5 % Coincidence error)
Maximum cabin pressure 0.6 MPa
Maximum cabin vacuum 0.08 MPa
Maximum viscosity detection
350 cps (40 sensor)
250 cps (32 sensor)
100 cps (25 sensor)
Ambient temperature 10 to 40°C
Interface RS232
Internal Dimension 430 × 320 × 590 mm
Internal Weight 55 kg
External Packaging Dimension 450 × 340 × 620 mm
External Packaging Weight 75 kg
Power 80 W
Power supply 220 V, 50 Hz

Features

Precision syringe sampling with stable sampling rate

Built-in threshold- diameter curve

8 Measurement channels

Wide Detection range

Large touch screen display

Large data processing, classification, verification

Built – in printer RS232 interface with connection to printer

Instrument calibration and verification as per international standards

FAQs

1. What are the key factors to consider when selecting this analytical equipment for laboratory use?

When selecting this analytical equipment, key factors include: 1) The instrument's precision and accuracy to meet experimental requirements; 2) Compatibility with the types of samples and applications you will be working with; 3) Ease of use and maintenance requirements; 4) Safety features to protect operators and samples; 5) Calibration and certification requirements; 6) Cost-effectiveness and long-term reliability; 7) Technical support and service availability from the supplier. Evaluating these factors ensures that the instrument meets your laboratory's specific needs.

2. How often should this analytical equipment be calibrated to ensure accurate results?

It is recommended to calibrate this analytical equipment at least once every 6-12 months under normal usage conditions. More frequent calibration (every 3-6 months) may be necessary if: 1) The instrument is used continuously for extended periods; 2) It is exposed to harsh environmental conditions (temperature fluctuations, humidity, dust); 3) It is used for critical applications requiring high precision; 4) The instrument has been moved or subjected to physical shock; 5) Calibration checks indicate significant deviations from standard values. Always follow the manufacturer's calibration guidelines.

3. What are the common maintenance procedures required for this analytical equipment?

Common maintenance procedures include: 1) Regular cleaning of the instrument's exterior and sample contact surfaces to prevent contamination and ensure proper operation; 2) Inspection of moving parts (if applicable) for wear and lubrication as recommended by the manufacturer; 3) Replacement of consumable parts (filters, bulbs, reagents) according to usage guidelines; 4) Verification of performance parameters (temperature, pressure, voltage) regularly; 5) Storage of the instrument in a clean, dry environment when not in use; 6) Keeping a maintenance log to track service activities and identify potential issues early; 7) Following the manufacturer's specific maintenance schedule for optimal performance and longevity.

4. Can this analytical equipment be used with different types of samples, and are there any compatibility considerations?

Yes, this analytical equipment can be used with a variety of sample types, but compatibility considerations are important: 1) Check if the instrument's materials are compatible with the chemical composition of your samples (avoid corrosive or reactive substances that may damage the instrument); 2) Ensure that the sample volume and concentration are within the instrument's recommended range; 3) For biological samples, consider sterility requirements and potential cross-contamination risks; 4) For volatile or hazardous samples, ensure proper ventilation and safety precautions are in place; 5) Consult the manufacturer's documentation for specific sample compatibility guidelines, and perform compatibility tests with new sample types before regular use.

5. What should be done if the analytical equipment malfunctions or produces inconsistent results?

If malfunctions or inconsistent results occur: 1) First, check the instrument's user manual for troubleshooting guidelines specific to the issue; 2) Verify that the instrument is properly calibrated and within its calibration period; 3) Inspect for obvious issues such as loose connections, damaged parts, or incorrect settings; 4) Clean the instrument thoroughly to remove any contamination that may affect results; 5) Repeat the experiment with a known standard sample to determine if the issue is with the instrument or the sample; 6) If the problem persists, discontinue use of the instrument to prevent further damage or inaccurate results; 7) Contact technical support or a qualified service technician for diagnosis and repair, providing details of the issue and any troubleshooting steps already performed.

Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.