Introduction
Veterinary Hematology Analyzer processes up to 50 samples per hour, making it suitable for high-volume veterinary facilities. It measures 21 hematology parameters and provides 3 diagnostic histograms. It features automatic maintenance and a fully enclosed reagent system. hematology analyzer features a 9.7-inch touchscreen display for easy and intuitive operation.
Applications
Veterinary Hematology Analyzer is used for detecting blood abnormalities, assessing health status, and monitoring treatment progress in animals. It is suitable for use in veterinary hospitals, pet clinics, zoos, animal research laboratories, drug research centers, quarantine departments, and livestock or animal husbandry institutions.
Specifications
| Throughput | 50 samples/hour |
| Principle | Impedance for WBC differentiation, WBC/RBC/PLT count & Colorimetric method for HGB |
| Channel | 2 |
| Measured Parameters (21) | MID#, GRAN#, LYM#, RBC, MCV, MCHC, RDW-CV, MPV, PCT, P-LCC, WBC, LYM%, MID%, GRAN%, HGB, HCT, MCH, RDW-SD, PLT, PDW, P-LCR |
| Histograms | 3 (WBC, RBC, PLT) |
| Sample Volume | Venous blood: 10 μL; Capillary blood: 10 μL; Prediluted: 20 μL |
| Calibration | Manual calibration, Automatic calibration, Fresh blood calibration |
| Counting Modes | Venous blood / Capillary blood / Prediluted blood |
| Printing Mode | Auto print / Manual print |
| LIS System | Supports Bi-directional LIS |
| Languages Supported | Chinese, English, Spanish, French |
| Reagents Used | Diluent 10L, Hemolytic Agent 250 mL (animal-specific), Probe Cleaner 50 mL |
| User Interface | 9.7-inch touchscreen, Mouse & Keyboard support (optional) |
| Output | Built-in Printer, Supports External Printers |
| Printer Paper Size | 57 × 30 mm |
| Communication Ports | 4× USB, 1× Network Port, 1× RS232 Serial Port |
| Storage Capacity | Stores 100,000 results with histograms |
| Data Transmission | HL7, Bi-directional LIS & multiple compatible protocols |
| Unclogging System | Reverse high-pressure flush |
| Working Temperature | 10 to 40 °C |
| Humidity Limit | ≤ 80% RH |
| Atmospheric Pressure Range | 86.0 to 106.0 kPa |
| Carry-over Rate | WBC ≤ 0.5×10⁹/L; RBC ≤ 0.05×10¹²/L; HGB ≤ 2 g/L; PLT ≤ 10×10⁹/L |
| Linearity Range | WBC: 1-100 ×10⁹/L; RBC: 1.0-7.0×10¹²/L; HGB: 20-200 g/L; PLT: 10-1000×10⁹/L |
| Precision | WBC ≤ 5%, RBC ≤ 2%, HGB ≤ 1.5%, PLT ≤ 10%, MCV ≤ 1.5%, MCHC ≤ 3%, RDW: 3-5%, HCT: 1.5-17.5% (CV%) |
| Power Supply | AC100-240V, 50/60Hz |
| Unit Dimensions (W×D×H) | 295 × 505 × 430 mm |
| Reagent Package Dimensions | 340 × 340 × 340 mm (Diluent), 200×150×100 mm (Hemolytic/Probe Cleaner) |
| Package Dimensions | 404 × 528 × 616 mm |
| Reagent Weight | 13 kg |
| Main Unit Net Weight | 20 kg; |
| Main Unit Gross Weight | 23 kg |
Features
Supports multiple species with 7 preset modes
Fully automatic detection with abnormal alerts
One-button maintenance with auto diagnostics
Enclosed reagent system to prevent contamination
Remote access and online QC supported
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.
Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.