Introduction
Horizontal is a 250 L capacity Class B, microprocessor controlled full automatic autoclave with vacuum drying function. It is equipped with advanced motorized or sliding door with an effective sterilization temperature of 139 °C. With safety features like auto door lock against raised safety pressure and automatic shutoff preventing overload of current this autoclave is an ideal laboratory device.
It uses high-temperature and high-pressure steam to achieve effective sterilization of laboratory equipment, media, and waste. The advanced control system ensures thorough sterilization while preventing overheating or pressure buildup.Applications
Horizontal autoclave is suitable for medical, laboratory, microbiological, clinical use and to sterile the waste before discarding.
Specifications
| Capacity | 250 L |
| Sterilization Temperature | 139 °C |
| Sterilization Pressure | 0.22MPa |
| Steam Pressure and the Consumption | 0.4 - 0.6 MPa / 15 kg |
| Water Source Pressure and the Consumption | 0.2 to 0.3 MPa / 150 kgs |
| Compressed Air Pressure | 0.6 - 0.8 MPa |
| Inner Chamber Dimension ( L × W × H ) | 800 × 600 × 600 mm |
| Outer Dimension ( L × W × H ) | 1100 × 1300 × 1700 mm |
| Net Weight | 650 kg |
| Power Supply | 26 kW |
Features
Autoclave offers single door design, it can be customized to optional double door design
Class B, fully automatic with vacuum drying
Digital display to monitor working status and parameters
Working pressure: 0.22 mPa
Working temperature: 139°C
Door lock feature: Door cannot be opened till the pressure inside the chamber doesn’t fall to 0.027 mPa
Safe door lock system ensuring safety / protection during sterilization
Automatic power cut off with alarm with low water levels indication
Motorized or sliding door with pneumatic sealing to prevent loss of sterile steam
Controlled power with built in steam generator
FAQs
1. Does the Horizontal Laboratory Autoclave notify when sterilization is complete?
Labron Horizontal Laboratory Autoclave is equipped with an end-of-cycle alarm system that alerts users when the sterilization and drying process is completed safely. The system automatically releases pressure and adjusts the temperature to safe levels, ensuring no manual intervention is required. Visual indicators on the digital display confirm when the entire cycle is finished and ready for unloading. This feature ensures that users can handle sterilized materials without delay and with complete safety, preventing overheating or pressure risks.
2. Can Horizontal Laboratory Autoclave sterilize liquids like culture media?
Horizontal Laboratory Autoclave efficiently sterilizes liquids such as culture media, nutrient broths, and other lab solutions with specially designed cycles to prevent spillage. it allows customizable temperature and pressure settings to ensure sensitive media do not boil over or degrade during sterilization. The system uses a vacuum release mechanism to safely depressurize without causing any liquid overflow or contamination. These features make the autoclave widely used in research and industrial labs for sterilizing various types of liquid preparations. Thus, it ensures high-quality and contamination-free media ready for laboratory use.
3. Can the Horizontal Laboratory Autoclave be customized with double doors?
Horizontal Laboratory Autoclave is equipped with a standard single door but can be customized with a double door on request. The double-door design is ideal for pass-through sterilization in clean rooms and sterile zones. It helps separate clean and unclean areas, preventing cross-contamination. This option is perfect for hospitals, pharmaceutical labs, and industries needing barrier systems. Double doors also improve workflow efficiency in sensitive environments. It ensures the safe and smooth transfer of materials between rooms.
4. Can the Horizontal Laboratory Autoclave be used continuously throughout the day?
Horizontal Laboratory Autoclave is built for heavy-duty and continuous use, allowing back-to-back sterilization cycles without losing performance. Its strong stainless-steel chamber and advanced cooling and pressure systems prevent overheating, even during long operations. Safety interlocks and alarms keep track of the autoclave’s performance to avoid accidents and ensure smooth operation. Energy-efficient features help reduce power use, even when running all day, saving on lab costs. This makes it a great choice for labs that need frequent and fast sterilization of equipment and materials.
5. How much power does the Horizontal Laboratory Autoclave consume during operation?
Horizontal Laboratory Autoclave is designed with a 26 kW power rating that supports rapid heating and efficient sterilization of large loads. The autoclave's heating elements are optimized for quick steam production, ensuring fast cycle completion without unnecessary energy waste. Advanced digital control systems help regulate power usage effectively, especially during standby or idle periods. By combining high performance with intelligent energy management, it delivers excellent results without excessive power bills. This makes it an ideal choice for labs seeking both speed and energy efficiency in sterilization processes.
6. Does the Horizontal Laboratory Autoclave require compressed air for operation?
Compressed air between 0.6 to 0.8 MPa is essential for operating key functions like pneumatic door sealing and vacuum drying in the It Horizontal Laboratory Autoclave. The use of compressed air ensures a completely airtight chamber, preventing leaks and maintaining optimal sterilization conditions inside. It also assists during the drying phase by facilitating effective moisture removal from sterilized materials, ensuring they are ready for immediate use. Connection ports are provided for easy integration with laboratory air supply systems, making setup and operation hassle-free.
Tips: For more details, quotes and customized solutions, please contact us anytime. We’re always ready to serve you professionally.