Introduction

Heating Incubator chamber has a volume of 1000L and can maintain temperatures from 10°C above room temperature to 250°C. This includes a controller with an over-temperature alarm and timing function for accurate temperature control. It is made from 304 stainless steel with a mirror-polished finish, providing a sleek appearance. The front damper adjustment of our heating incubator optimizes gas convection for improved performance. It provides a stable and controlled environment for cell culture, microbial growth, and other biological experiments. The precise temperature and humidity control ensure consistent experimental conditions and reliable results.

Applications

The Heating Incubator is designed for desiccation, torrefaction, wax-melting, and sterilization in mining, laboratories, and research institutes.

Specifications

Temperature Range RT +10℃ to 250℃
Chamber Volume 1000 L
Shelves 4
Display Resolution 0.1 ℃
Temperature Stability ±1 ℃
Temperature uniformity ±3%
Ambient Temperature +5℃ to 40℃
Timing Range 0 to 9999 min
Power consumption 5200 W
Power Supply 380V 50Hz
Interior Dimension 800 × 700 × 1590 mm
Exterior Dimension 1050 × 880 × 206 mm
Packaging Dimension 1270 × 950 × 2280 mm
Net Weight 270 Kg
Gross Weight 377 Kg

Features

Advanced air circulation system

Silicon door gasket for effective sealing

Easy to clean and maintain

Optional Independent over-temperature alarm

FAQs

1. How to calibrate the temperature and humidity controls in the laboratory incubator?

Calibration should be performed by qualified personnel using calibrated temperature and humidity sensors: 1) Place the calibrated sensors at different locations inside the incubator chamber to ensure uniform distribution; 2) Set the incubator to the desired temperature and humidity levels and allow it to stabilize for at least 24 hours; 3) Compare the readings from the incubator's display with the calibrated sensors; 4) Adjust the controls according to the manufacturer's instructions if discrepancies are found; 5) Document the calibration results and schedule regular recalibration (every 6-12 months).

2. What is the optimal way to arrange samples inside the incubator for uniform conditions?

To ensure uniform conditions: 1) Avoid overcrowding the incubator, leaving sufficient space between samples for air circulation; 2) Place samples away from the chamber walls, heating/cooling elements, and door to prevent temperature fluctuations; 3) Use shelves or racks that allow air to circulate freely around all samples; 4) Avoid placing large or heat-generating equipment inside the incubator; 5) Rotate samples periodically if long-term incubation is required to account for any minor temperature variations.

3. Can the incubator be used for anaerobic or CO2-enriched environments?

Specialized incubators (anaerobic incubators, CO2 incubators) are designed for these environments. Standard incubators may not maintain the required gas concentrations or oxygen-free conditions. For anaerobic applications, use an incubator equipped with a gas control system that maintains oxygen levels below 1% and may include catalysts to remove oxygen. For CO2-enriched environments, use a CO2 incubator that maintains precise CO2 concentrations (typically 5-10%) and humidity levels, essential for cell culture applications.

4. How to clean and disinfect the incubator to prevent contamination?

Regular cleaning and disinfection steps: 1) Turn off the incubator and allow it to cool down to room temperature; 2) Remove all samples and shelves; 3) Wipe the interior surfaces with a mild detergent solution, followed by a disinfectant compatible with the incubator materials (such as 70% ethanol); 4) For stubborn contamination, use a specialized laboratory disinfectant according to the manufacturer's instructions; 5) Clean the shelves and any accessories separately; 6) Allow all surfaces to dry completely before reassembling and restarting the incubator; 7) Perform this procedure monthly or more frequently if contamination is suspected.

5. What should be done if the incubator temperature fluctuates significantly?

If temperature fluctuations occur: 1) Check if the incubator door is properly closed and sealed; 2) Ensure the incubator is not placed near heat sources, air conditioning vents, or drafty areas; 3) Verify that the incubator is not overloaded, which can restrict air circulation; 4) Check the air filter (if equipped) and replace it if dirty; 5) Inspect the heating and cooling elements for proper operation; 6) If the problem persists, contact technical support for further diagnosis and repair, as it may indicate a malfunction in the temperature control system.