Introduction

Shaking Incubator operates on a forced convection circulation system with a capacity of 580L. It supports a wide range of cyclotron frequencies and a temperature range from RT+5℃ to 65℃. It has a track drive mode for smooth and reliable operation. Controlled by a microcomputer, it offers various modes like fixed value, static, and multi-segment program control. Our incubator features a built-in observation window for clear visibility of chamber contents. It provides a stable and controlled environment for cell culture, microbial growth, and other biological experiments. The precise temperature and humidity control ensure consistent experimental conditions and reliable results.

Applications

Our Shaking Incubator is used in laboratories for microbial culture, cell cultivation, and biochemical reactions. They ensure precise temperature control and consistent agitation for optimal growth and experimental conditions.

Specifications

Capacity 580 L
Circulation Forced convection
Drive Mode Track drive
Shake Mode Reciprocate
Orbit Diameter Φ 0 to 50
Cyclotron Frequency Range 0 (for static culture); 20 to 260 rpm/min
Cyclotron Frequency Resolution ±1
Temperature Range RT+5℃ to 65℃
Temperature Fluctuation ≤0.1 (37℃)
Temperature Resolution ± 0.1
Temperature Uniformity ±0.5 (37℃)
Flask Clamp Max. Capacity 100ml×140 or 250ml×90 or 500ml×72 or 1000ml×42
Flask Clamp Standard Capacity 250ml×45, 500ml×3
Controller microcomputer
Control Mode Fixed value mode, static mode, multi-segment program control mode
Display Large high-definition LC
Timer 0 to 9999 h (or continuous operation)
Safety Device Self-diagnosis, visible & audible over temperature, abnormal status alarm
Power Consumption 1100 W
Power Supply AC 220 / 110V, 50 / 60Hz
Platform Dimension (W×D×H) 950×550 mm
External Dimension (W×D×H) 1200 × 800 × 1630 mm
Packaging Dimension (W×D×H) 1355 × 950 × 1795 mm
Net Weight 500 Kg
Gross Weight 580 Kg

Features

Internal circulation system

Intelligent control

Dedicated sheet metal processing

Anti-scalding designed door handles

Visible and audible alarm system

Fault self-diagnosis function

Auto-stop operation when door opened

Automatically locks when speed is out of control

FAQs

1. How to calibrate the temperature and humidity controls in the laboratory incubator?

Calibration should be performed by qualified personnel using calibrated temperature and humidity sensors: 1) Place the calibrated sensors at different locations inside the incubator chamber to ensure uniform distribution; 2) Set the incubator to the desired temperature and humidity levels and allow it to stabilize for at least 24 hours; 3) Compare the readings from the incubator's display with the calibrated sensors; 4) Adjust the controls according to the manufacturer's instructions if discrepancies are found; 5) Document the calibration results and schedule regular recalibration (every 6-12 months).

2. What is the optimal way to arrange samples inside the incubator for uniform conditions?

To ensure uniform conditions: 1) Avoid overcrowding the incubator, leaving sufficient space between samples for air circulation; 2) Place samples away from the chamber walls, heating/cooling elements, and door to prevent temperature fluctuations; 3) Use shelves or racks that allow air to circulate freely around all samples; 4) Avoid placing large or heat-generating equipment inside the incubator; 5) Rotate samples periodically if long-term incubation is required to account for any minor temperature variations.

3. Can the incubator be used for anaerobic or CO2-enriched environments?

Specialized incubators (anaerobic incubators, CO2 incubators) are designed for these environments. Standard incubators may not maintain the required gas concentrations or oxygen-free conditions. For anaerobic applications, use an incubator equipped with a gas control system that maintains oxygen levels below 1% and may include catalysts to remove oxygen. For CO2-enriched environments, use a CO2 incubator that maintains precise CO2 concentrations (typically 5-10%) and humidity levels, essential for cell culture applications.

4. How to clean and disinfect the incubator to prevent contamination?

Regular cleaning and disinfection steps: 1) Turn off the incubator and allow it to cool down to room temperature; 2) Remove all samples and shelves; 3) Wipe the interior surfaces with a mild detergent solution, followed by a disinfectant compatible with the incubator materials (such as 70% ethanol); 4) For stubborn contamination, use a specialized laboratory disinfectant according to the manufacturer's instructions; 5) Clean the shelves and any accessories separately; 6) Allow all surfaces to dry completely before reassembling and restarting the incubator; 7) Perform this procedure monthly or more frequently if contamination is suspected.

5. What should be done if the incubator temperature fluctuates significantly?

If temperature fluctuations occur: 1) Check if the incubator door is properly closed and sealed; 2) Ensure the incubator is not placed near heat sources, air conditioning vents, or drafty areas; 3) Verify that the incubator is not overloaded, which can restrict air circulation; 4) Check the air filter (if equipped) and replace it if dirty; 5) Inspect the heating and cooling elements for proper operation; 6) If the problem persists, contact technical support for further diagnosis and repair, as it may indicate a malfunction in the temperature control system.