Introduction
Class II Biosafety Cabinet Type B2 a maximum front window opening of 420 mm for ease of access and flexibility. It is designed with an inflow velocity of 0.53 m/s to ensure effective containment. cabinet features a programmable time reserve function, allowing users to schedule operations for enhanced workflow efficiency. Its LCDs display real-time data on airflow, filter pressure, usage time, UV status, temperature, and humidity.
Applications
Class II Biosafety Cabinet Type B2 is widely used in Clinical and Diagnostic Labs, Microbiology and Virology Labs, Pharmaceutical and Biotechnology R&D, Hospitals and Healthcare Facilities, Universities and Research Institutions, Public Health and Government Labs.
Specifications
| Maximum Opening | 420 mm (17 inches) |
| Tested Opening | Safety Height 200mm (8 inches) |
| Work Surface Height | 750 mm (customizable) |
| Air Inflow Volume | 0.53±0.025 m/s |
| Downflow Velocity | 0.33±0.025 m/s |
| ULPA Filter | Two, 99.9995% efficiency @ 0.12μm, with filter life indicator |
| Front Window | Motorized, two-layer laminated toughened glass ≥ 6mm, anti-UV |
| Noise | Less than 67dB |
| UV Lamp | 30W×1 |
| UV Lamp Note | UV timer, UV life indicator, emission of 253.7 nm for decontamination |
| Illuminating Lamp | LED lamp |
| Lamp Power | 12W×2 |
| Illumination | Greater than 1000 Lux |
| Power Consumption | 850W |
| Waterproof Sockets | Two, total load: 500W |
| Control System | Microprocessor |
| Airflow System | 0% air recirculation, 100% air exhaust |
| Alarm | Abnormal airflow, filter replacement, unsafe window height, and high pressure |
| Exhaust Duct | 4 meters PVC Duct, Diameter: 300 mm |
| Work Zone Material | 304 stainless steel |
| Main Body Material | Cold-rolled steel, anti-bacterial powder coating |
| Caster | Footmaster caster |
| Power Supply | 220V±10%, 50Hz |
| Main Body Dimension | 1460×1000×1820 mm |
| Exhaust Blower Dimension | 970×810×630 mm |
| External Dimension (WDH) | 1300×750×2250 mm |
| Internal Dimension (WDH) | 1150×600×660 mm |
| Weight | 280 kg |
Features
Motorized front window
Auto-adjusting air speed
Power failure memory recovery
Remote control access
Redundant Safety Alarms
Foot Switch Control
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.