Introduction
Class II Biosafety Cabinet Type B2 designed with an air inflow velocity of 0.53 m/s, ensuring optimal containment and protection. It features a tested safety working height of 200 mm and a maximum front window opening of 420 mm. cabinet features a convenient remote control for full function access. It also includes an intuitive LCD for clear real-time monitoring.
Applications
Class II Biosafety Cabinet Type B2 is widely used in Clinical and Diagnostic Labs, Microbiology and Virology Labs, Pharmaceutical and Biotechnology R&D, Hospitals and Healthcare Facilities, Universities and Research Institutions, Public Health and Government Labs.
Specifications
| Maximum Opening | 420 mm (17 inches) |
| Tested Opening | Safety Height 200mm (8 inches) |
| Work Surface Height | 750 mm (customizable) |
| Air Inflow Volume | 0.53±0.025 m/s |
| Downflow Velocity | 0.33±0.025 m/s |
| ULPA Filter | Two, 99.9995% efficiency @ 0.12μm, with filter life indicator |
| Front Window | Motorized, two-layer laminated toughened glass ≥ 6mm, anti-UV |
| Noise | Less than 67dB |
| UV Lamp | 30W×1 |
| UV Lamp Note | UV timer, UV life indicator, emission of 253.7 nm for decontamination |
| Illuminating Lamp | LED lamp |
| Lamp Power | 12W×2 |
| Illumination | Greater than 1000 Lux |
| Power Consumption | 800W |
| Waterproof Sockets | Two, total load: 500W |
| Control System | Microprocessor |
| Airflow System | 0% air recirculation, 100% air exhaust |
| Alarm | Abnormal airflow, filter replacement, unsafe window height, and high pressure |
| Exhaust Duct | 4 meters PVC duct, Diameter: 300 mm |
| Work Zone Material | 304 stainless steel |
| Main Body Material | Cold-rolled steel, anti-bacterial powder coating |
| Caster | Footmaster caster |
| Power Supply | 220V±10%, 50Hz |
| Main Body Dimension | 1220×1000×1820 mm |
| Exhaust Blower Dimension | 970×810×630 mm |
| External Dimension (WDH) | 1100×750×2250 mm |
| Internal Dimension (WDH) | 940×600×660 mm |
| Weight | 240 kg |
Features
Foot switch control
Negative pressure work area
Audio/visual alarms
Memory function
ULPA filter and UV lamp life indicators
FAQ
Q1: What is the main application of this laboratory equipment?
A1: It is mainly used for general laboratory sample processing in laboratory settings, such as sample preparation, analysis, or processing.
Q2: Does the product come with a warranty?
A2: Yes, it provides a 1-year warranty for the main unit (excluding consumable parts), covering manufacturing defects and quality issues.
Q3: Is installation required after receiving the product?
A3: No, it is ready to use after unpacking. Simply connect the power supply (confirm the voltage matches) and follow the operation manual.
Q4: What are the power supply requirements?
A4: It supports AC 220V/50Hz or 110V/60Hz (selectable based on region), please confirm the voltage specification before use.
Q5: How to get technical support if there is a problem?
A5: You can contact our technical support team via email or phone. We provide online troubleshooting guidance and on-site service if needed.